Electrical and Mechanical

CPU

i7, 2.53 GHz

Memory

4 GB

Hard Disk

8 GB Solid State Disk (up to 64GB optional). A removable solid-state disk for classified data storage is also optionally available

Interfaces

RS-422, RS-232, USB, Ethernet, VGA

Display

Optional sunlight-readable LCD sVGA display 600x800 resolution or a Tablet PC

Control

Optional touch-screen (part of display)

Input Power

One or three phase
105 VAC to 250 VAC
45 Hz to 420 Hz
800 W (max) input power

DC Input power (Optional)

28 VDC (750 W max input power)

System Power (PXI Backplane)

+5 VDC @ 35 A
+3.3 VDC @ 35 A
+12 VDC @ 3 A
-12 VDC @ 3 A
+19 VDC @ 5 A (for Tablet)

Output Power (Auxiliary)

28 VDC @ 1 A
5 VDC @ 1 A

UUT Output Power (Optional)

One DC/DC power supply brick, 240 W max, single or triple output

Power Protection

Over / Under Voltage
Over Current
Reverse Input DC Voltage
Over / Under Temperature

Environmental*

Operating Temperature

-40 ºC to +63 ºC

Storage Temperature

-44 ºC to +71 ºC

Relative Humidity

95%, non-condensing

Altitude

15,000 ft (operating)
30,000 ft (non-operating)

Loose Cargo Vibration

IAW MIL-STD-810F, Method 514, Procedure II

Vibration (Sine)

IAW MIL-T-28800
Paragraph 4.5.5.3.2

Vibration (Random)

IAW MIL-T-28800
Paragraph 4.5.5.3.1

Shock (Half-Sine)

IAW MIL-T-28800
Paragraph 4.5.5.4.1
7G Operating
30G Non-Operating


Salt Fog

MIL-STD-810F, Method 509, Procedure I

Dust

MIL-STD-810F, Method 510, Procedure I

Rain

MIL-STD-810, Method 506, Procedure II

EMI (Optional)

MIL-STD-461E

Cooling

13 fans, 350CFM (max)
(temperature based speed control)

Heating

Built-in 345 W heater

Size

25" L x 19" W x 20" D

Weight

60 lb - basic; varies by configuration

System Monitoring and Control - Smart Features*

Control

A dedicated microcontroller on the Power Board is capable of reading back measurements and controlling user I/O

Temperature Monitoring

2 sensors on Power Board (with readback)
8 sensors on the PXI backplane

Voltage Monitoring

PXI Power: +3.3 V, +5 V, +12 V, -12 V
+28 V (main), user DC / DC power supply

Current Monitoring

PXI Power: +3.3 V, +5 V
Rectified AC current, Fan current, user DC / DC power supply

Frequency Monitoring

AC Line Frequency (low resolution)

Measurement Resolution

8-bit

Relay Control

One optional 25 A 3PDT relay (socketed)
Two optional 10 A DPDT relay (socketed)

User I/O

32-bit, factory configurable as TTL or open-collector
4 groups of 8 with user direction control

*Some environmental specifications require optional features such as heaters and EMI shielding. Note: Specifications are subject to change without notice.

Dynamic Digital Subsystem
Configuration 6U PXI format, (7) 32 channel DIO cards, (1) 16 channel DIO card,
programmable drive / sense per pin, 50 MHz vector rate (max)
Number of Hyrid Channels 224
Test Modes Dynamic or static
Data Output Formats (per channel) Drive Hi, Drive Lo, Hi-Z
Formatted Data: No return, Return to 1, Return to 0, Return to Hi-Z, Return to complement, Surround by complement; selectable on a per channel basis
Drive Data Timing (per channel) Data assert / de-assert based on Phases 1-4
Timing Sets 256 Timing Set groups with 4 Phases, 4 Windows, and 4K sequence steps)
Phase Programming Range (Assert/Return) 0 ns to 64 us (using the 500 MHz master clock)
Window Programming Range (Open/Close) 0 ns to 64 us (using the 500 MHz master clock)
Phase and Window Timing Resolution 1 ns, using the 500 MHz master clock
Capture Modes (per channel) Mask
Opening edge of Window
Closing edge of Window
Window – data is valid for entire window duration
Drive/Expect Mode Output: Drive Hi, Drive Lo, Hi-Z
Expect: 1, 0, OK, between states, or mask
Keep last
Toggle last
Accumulate CRC-16
Error Address Record Record address for memory errors
1K deep error memory
Drive Level Range -14 V to +26 V
Drive Voltage Level Range Min: 0.5 V p-p
Max:  26 V p-p
Drive Voltage Accuracy ± 25 mv, < 26 V p-p drive voltage
Sense Level Range -16 V to +22 V
Maximum Sense Voltage Range 26 V
Sense Voltage Resolution 16 bits
Pull-Up/Pull-Down Current Source/Sink ± 24 ma, programmable on a per channel basis
V commutate: -16 to +22 V., programmable on a per channel basis
Pull-Up/Pull-Down Current Source/Sink Accuracy ± 250 uA
Pull-Up/Pull-Down Current Source/Sink Resolution 16 bits
Trigger/Clock Signals (4) Aux signals for external triggering, clocking of digital subsystem, accessible via the receiver interface and the IIP
Mainframe Specifications
Input AC Power 120 / 208 VAC nominal, 3 phase, 5-wire wye configuration
20 A per phase, 47 to 63 Hz,
Input voltage range: 100 / 170 to 155 / 264 VAC
HV Input AC Power Configuration 230 / 400 VAC nominal, 3 phase, 5-wire wye configuration
10 A per phase, 47 to 63 Hz,
input voltage range: 223 / 380 to 310 / 528 VAC
Total Available DC Power PXI voltages: 800 W
Pin electronics: 4.4 KW
PXI Power
+5 V
+3.3 V
+12 V
-12 V
Note: Total output power cannot exceed 800 W.
60 A (max)
60 A (max)
25 A (max)
5 A (max)
Pin Electronics (via P5 connector)
VCC
VEE
3.3 V
+2 V to +30 V, @ 50 A
-4 V to -18 V, @ 80A
3.3 V @ 60 A
Power Supply Monitoring Monitored voltages: 3.3, 5, +12, -12, VIO value
Accuracy: ± 2% of reading
Temperature Monitoring Per slot monitoring, 1 reading/sec/slot
4 second moving average value
User selectable alarm criteria:
  • Maximum slot temperature
  • Average slot temperature
Accuracy: ± 2 °C
Default warning and shutdown limits: +50 °C & +70 °C
Warning and shutdown limits programmable via software driver
Status: Query via software driver and audible alarm for a warning limit condition
PXI Triggers Slots: 2 - 20
Number: 8 per segment
Software controlled segment mapping supports:
  • Isolate a trigger line within a segment
  • Map a trigger line left to right
  • Map a trigger line right to left
PXI Clock Integrated 10 MHz PXI clock with auto-detect function. Presence of an external 10 MHz PXI clock will disable the internal clock. PXI clock is distributed to all peripheral slots.
  • 10 MHz PXI clock accuracy: ±100 ppm
  • External input: Rear panel (TTL compatible) or via timing slot
  • Output: Rear panel, (TTL compatible)
Slots 20 PXI or cPCI Slots (19 instruments max)
Weight (core system) 100 lbs
Overall Dimensions 10U (17.5") H x 19" W x 24" D
Cooling Eight 100 CFM fans for system cooling. Integrated temperature monitoring via an on-board microcontroller with audible and software notification when preset temperature limits are exceeded. Fan speed control and monitoring is automatic and can be controlled / monitored via the GxChassis software.
Environmental and Compliance
Operating 0 °C to +50 °C
Storage -20 °C to +60 °C
Relative Humidity (operating) 5% to 80% RH,non-condensing
Relative Humidity (non-operating) 5% to 95% RH,non-condensing
Altitude (operating) Up to 2000 M
CE Compliance EN61010-1 (pending)
EN61326 (pending)

Note: Specifications are subject to change without notice.

介绍 >>   规格 >>   订货 >>
 
虹科 > 电子测试 > MTS 测试测量产品和解决方案
测试产品和解决方案
航空航天测试系统
生产制造测试系统
半导体测试系统
硬件
软件
PXI 仪器

公司
招聘人才
关于虹科
公司新闻

加关注


关于虹科
:
虹科积极进取,不断探索科技新领域。最优秀的员工和最好的技术能够接受最困难的挑战。为您找到最佳解决方案。虹科靠口碑赢得客户。

Email 新浪微博 论坛 点击发送消息给我

北京:010-5781 5040 | 上海:021-6728 2707 | 西安:029-8187 3816 | 广州:400 999 3848 | 成都 028-6138 2617
沈阳 024-8376 9335 | 深圳 0755-2267 7479 | 武汉 027-8193 9100 | 香港 6749 9159 | 伦敦 74 3120 4210

sales@dgruijun.com | 广州虹科电子科技有限公司 | 版权所有 | 广州科学城润慧科技园C栋6层

友情链接: